Harmonics 分析功能规格
1-50阶电压电流失真百分比V[n%]、A[n%]
1-50阶瓦特W[n]
1-50阶瓦特失真百分比W[n%]
1-50阶电压与电流角度DEG[n]
Vrms,Irms,Watt, PF
规格
7110-10K-HARM
7120-100K-HARM
量测数据
500Hz -3db 基于Butterworth 之数位滤波器
50Hz -0.03% reading, 60Hz -0.05% reading
量程
RMS/MEAN模式电压&电流精准度(23 ℃± 5 ℃ )
F单位为1KHz
当L-FILTER 设定ON: 45Hz~66Hz频段容许误差-0.03 %~-0.05 of reading
AC量测时基频频率超过200Hz时需将F-Filter关掉以量得*准确读值
**超过10KHz 以上频段7120才有支援
DC模式电压&电流精准度(23 ℃± 5 ℃ )
Power (W) 精准度(23℃±5℃)
AC power ranges (Auto or Manual)(40 ranges)范围up to 16KW *大Power(W)值依电压档位*高量程而定
DC ±0.2% reading ± 0.5% of range
15 Hz≦f
45 Hz≦f≦66 Hz ±(0.1% of reading + 0.1 % of range)
66 Hz<f≦1 kHz ±(0.2% of reading + 0.2 % of range)
1 kHz<f≦10 kHz ±(0.4% of reading + 0.3 % of range)± [{0.06×(f)}% of reading]
10 kHz<f≦100 kHz ±(0.5 % of reading + 0.5 % of range)±[{0.09×(f–10)}% of reading]
频率介于45 to 66 Hz: Add 0.3% of reading. 频率于45 to 66 Hz:范围以外Add 1% of reading
当功率因素PF 为0时,Watt的误差范围为
情况1: for 45 Hz<f
情况2: for 45 Hz>f 或f >66 Hz
Add ±{(3.5 + 0.5 × f)% of VA} for up to 100 kHz as reference data
The unit for frequency f is kHz.
当功率因素0 < PF ≦ 误差范围
When 0<PF≦1 (θ: phase angle of the voltage and current)
for 45 Hz≦f≦66 Hz. Add ± power reading *{tan(θ)*(0.5)}%
for f < 45 Hz, f >66 Hz. Add ± power reading *{ tanθ* (0.5×f+0.2) }%
参数量测范围
应用范围
量测应用于环保监视器、节能灯、交换式电源供应器、变压器、不断电系统、电动工具、3C电子产品、家电产品等
蓉公网安备 51012402000290号